V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Image Display Error/Artifact (1304); Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 16mar2020.
 
Event Description
It was reported that the unit's display was flashing and the navigation ring would not move.There was no patient involvement.
 
Manufacturer Narrative
G4: 07may2020.B4: 08may2020.The manufacturer's field service engineer (fse) performed troubleshooting at the customer's site.The fse replaced the navigation ring/front bezel and the issue was resolved.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
G4: 12feb2021.D4: udi#: (b)(4).H11: g5: k102985.H10 the replaced front bezel was returned for failure analysis.It was determined to be due to liquid ingress caused the reported navigation ring failure.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key9838316
MDR Text Key186020715
Report Number2031642-2020-00943
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/27/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/25/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-