BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/35; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 391018
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/08/2019
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment of a severely calcified lesion (70 percent stenosis degree) in a moderate tortuous proximal lad.Despite pre-dilatation the lesion could not be crossed with the device.It was detected that a stent deformation has occurred.
 
Additional Manufacturer Narrative
Only the stent was returned for investigation and subjected to a technical analysis.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that the stent is severely deformed, particularly stretched in one half of the stent.The delivery system was not returned for analysis.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.It should be noted that the ifu warns against the application of excessive force during lesion crossing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.5/35
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key9838834
Report Number1028232-2020-01298
Device Sequence Number2929156
Product Code NIQ
UDI-Device Identifier07640130439399
UDI-Public07640130439399
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/21/2020
Device Model Number391018
Device Catalogue NumberSEE MODEL NO.
Device Lot Number10181454
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2020
Initial Date Received by Manufacturer 03/12/2020
Supplement Date Received by Manufacturer05/15/2020
Initial Report FDA Received Date03/17/2020
Supplement Report FDA Received Date05/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-