| Model Number |
P10.7-80C245 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for evaluation.It is unknown if patient complied with post-operative physical restrictions.No root cause can be confirmed at this time.
|
| |
|
Event or Problem Description
|
|
Information was received that a revision procedure was performed on (b)(6) 2019.As per reporter, the nail failed to lengthen.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection of the returned nail showed score marks present on the outer surface of the nail.X-ray images of the internal components showed no damage and revealed the nail to be partially distracted.The nail was functionally tested and was able to distract and retract with the high speed magnet tool.The device met all specifications and the reported defect was unable to be duplicated or confirmed.
|
| |
|
Event or Problem Description
|
|
See updated fields d10, h3, h6, h10.
|
| |
|