| Model Number |
M00567340 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation results: a visual examination of the returned complaint device found that the balloon did not have any visual defects and was in a good condition.The catheter was kinked in several sections.The outer diameter of the catheter was measured in three sections, and all sections were found to be within specifications.Functional evaluation was performed and the balloon was inflated without problem; however, the balloon would not hold pressure due to a pinhole located over the proximal section of the ro marker.This failure is likely due to factors or conditions related to the procedure during the use of the device, such as the interaction between the scope and device, that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a maxforce biliary dilatation balloon was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the balloon was attempted to be inflated; however, it did not inflate.Reportedly, the balloon was then removed from the patient and was checked.A hole was noticed in the balloon.The procedure was completed with another maxforce biliary dilatation balloon.There were no patient complications reported as a result of this event.Investigation results revealed that the balloon had a pinhole; therefore, this is now an mdr reportable event.
|
| |
|
Search Alerts/Recalls
|
|
|