The excor blood pump,s/n (b)(4), was in use by the patient from (b)(6) 2020 until (b)(6) 2020 (46 days).We have reviewed the production records of the excor blood pump, s/n (b)(4).This pump was produced according to our specification.The blood pump is designed with a triple layer membrane separating the air chamber from blood chamber for safety reasons.The entire membrane consists of an air-side layer, a middle layer and a blood-side layer.In case of disruption in one of the triple layers, there are two more layers that will maintain the integrity of the air and blood chambers.The pump was returned to the manufacturer on 3/13/2020.A detailed investigation report will be provided as soon as it is available.
|
During initial visual examination of the returned blood pump, no defect could be detected.No graphite was seen.The blood pump was then tested for functional performance, where it did reach its required functional performance.The blood pump was fully filling and emptying.The pump was then disassembled for further testing and the membrane layers were individually examined.All three membrane layers were found to be intact.No membrane defect was found.The cause of the incomplete filling the clinic described could not be attributed to any defect of the blood pump.
|