MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD VIO 36IN 1 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD VIO 36IN 1 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J371H
Medical Device Problem Codes Material Frayed (1262); Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/21/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device pmz887 batch number, and no non-conformances were identified.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that the patient underwent an unknown procedure on (b)(6) 2020 and suture was used.During the procedure, the suture frayed at the swage and detached from the needle at the swage.A like device was used to complete the procedure with no delay in the surgery.There were no adverse patient consequences reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 04/14/2020.Additional information: investigation summary: it was reported performance pull off - suture needle and performance fraying suture intra-op.Unopened samples were returned for analysis.The complaint sample was not received.During the visual inspection of six unopened samples, no defects were found on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, fraying or damages were observed.A functional test was performed and the pull force results were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the samples received, no performance pull off - suture needle or performance fraying suture was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 36IN 1 S/A CTX
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key9842175
Report Number2210968-2020-02145
Device Sequence Number19492200
Product Code GAM
UDI-Device Identifier10705031039391
UDI-Public10705031039391
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberJ371H
Device Catalogue NumberJ371H
Device Lot NumberPMZ887
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Initial Date Received by Manufacturer 02/21/2020
Supplement Date Received by Manufacturer03/19/2020
Initial Report FDA Received Date03/17/2020
Supplement Report FDA Received Date04/14/2020
Patient Sequence Number1
-
-