• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number NEU_ENTERRA_INS
Device Problems Migration or Expulsion of Device (1395); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bruise/Contusion (1754); Fever (1858); Unspecified Infection (1930); Pain (1994); Seroma (2069)
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer (con) regarding a patient who was implanted with a neurostimulator for gastric stimulation it was reported that an enterra patient who has an infection around the pocket site was treated with antibiotics.No further complications noted or anticipated.
 
Manufacturer Narrative
Continuation of d11: product id neu_unknown_lead serial# unknown product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from a manufacturer representative (rep).It was reported that the rep didn¿t have the device and they weren¿t sure if it was thrown away or not.
 
Manufacturer Narrative
Section d information references the main component of the system.Other relevant device(s) are: product id: neu_unknown_lead; product type: lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Additional information was received and the manufacturer representative stated that the patient had a lot of pain after surgery and presented with a mild seroma as well as bruising at her port sites the following day.  patient came to the doctor's office on (b)(6) 2020 with intense pain and a low grade fever (100 degrees).  on (b)(6) he pulled 15 cc¿s of thin bloody fluid from her pocket site.  they cultured the fluid but no growth was found.  she presented to the er with pain on a couple of occasions where they did a ct scan and noted a small amount of fluid at the pocket site.  at some point she was placed on antibiotics as well as an antifungal.  she seem to do well on the antifungal for a small period of time.  they spoke to her via phone on (b)(6) and she was having intense pain but was having good response to her nausea and vomiting symptoms.  on (b)(6) they saw patient in the office and decided to remove her device due to continued pain.  on (b)(6) they removed it and noted a little hematoma at the pocket site as well as leads that were intertwined with some adhesions.  the health care professional still has no clue as why she was experiencing the severe pain.No further complications noted or anticipated.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
MDR Report Key9842892
MDR Text Key183759298
Report Number2182207-2020-00014
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 04/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberNEU_ENTERRA_INS
Device Catalogue NumberNEU_ENTERRA_INS
Was Device Available for Evaluation? No
Date Manufacturer Received04/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
Patient Weight70
-
-