This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 17, 2020.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date); g4 (date received by manufacturer); g7 (indication that this is a follow-up report); h2 (follow-up due to additional information and device evaluation); h3 (device evaluated by manufacturer); h4 (device manufacture date); h6 (identification of evaluation codes 10, 11, 3331, 213, 67).Method code #1: 10 - testing of actual/suspected device.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 - analysis of production records.Results code: 213 - no device problem found.Conclusions code: 67 - no problem detected.Visual inspection was performed on the returned sample.There was blood noted within the valve, no other visual anomalies were noted on the sample.Ops valves are subject to a 100% leak test, which includes five different tests the units are run through to ensure there are no leaks and both the umbrellas and duckbills are functioning properly.The returned sample was setup to be manually run through each of these tests to determine if the umbrellas were functioning properly and not allowing air to enter the valve.The returned sample met specification and functioned as intended.The retention sample was functioned as intended.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|