• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ASPHERE M SPEC 11/13 36 +0; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ASPHERE M SPEC 11/13 36 +0; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-31-100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Edema (1820); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Test Result (2695); No Code Available (3191)
Date of Event 06/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pcf and medical records received.After review of medical records, patient was revised to address left hip modular metal-on-metal total hip arthroplasty with adverse local tissue reaction and possible metal toxicity.Patient has developed left hip pain and mild abductor irritation but no severe weakness.Laboratory studies showed that patient had elevated serum metal ion levels with cobalt greater than 8 and chromium greater than 4.Operative notes stated that upon entry into the joint, there was expression of only a very scant amount of clear, benign-appearing straw-colored joint fluid.There were a few, patchy areas of dark gray to black stained synovium but the majority of the periprosthetic synovial soft tissues were noted to be grossly healthy appearing without granulation, friability, or necrosis.There was no gross evidence of infection.Doi: (b)(6) 2010.Dor: (b)(6) 2019 (left hip).Patient was bilateral.See (b)(4) for the right hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPHERE M SPEC 11/13 36 +0
Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9843411
Report Number1818910-2020-08313
Device Sequence Number16328781
Product Code JDI
UDI-Device Identifier10603295033370
UDI-Public10603295033370
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K082585
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2019
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2015
Device Model Number1365-31-100
Device Catalogue Number136531100
Device Lot Number3106522
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/28/2020
Supplement Date Received by Manufacturer04/17/2020
Initial Report FDA Received Date03/17/2020
Supplement Report FDA Received Date04/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-