| Model Number |
10492730 |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This issue has been escalated to software engineering for evaluation.The trace log provided by the customer had the incorrect time frame.Siemens has requested the trace log with the correct time frame.The cause of this event is unknown.
|
| |
|
Event or Problem Description
|
|
The customer reported they ran a patient on the rp 500 and the user confirms the patient id was scanned correctly, all results on the patient's chart are correct, the printouts from the analyzer are correct and the rapidcomm results are correct.When the user recalled the patient list, several results from the previous 24 hours were all overwritten with one, different patient id.There was no report of injury due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens has made multiple "attempts" to acquire additional information.There has been no response from the customer.Since the hard drive is not available, no conclusion can be made.The customer stated that all rapidcomm results are correct, all printouts are correct and all patient results on the chart are correct.The logs provided do not include information required to determine the root cause.
|
| |
|