| Model Number |
MMT-722NAS |
| Medical Device Problem Codes |
Failure to Cycle (1142); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm at the time of prime rewind.Customer had unspecified alarm.The customer stated that they were unable to clear the alarm.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform the a21 error test, prime or a33 test, excessive no delivery test and occlusion test or verify a33 alarm due to motor error alarm.Device received with normal operating current.Device passed self test.Device passed off no power test.No unexpected e/a alarm anomalies noted.
|
| |
|
Search Alerts/Recalls
|