| Model Number |
8888135132 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led a photographic evaluation of one device.The visual inspection of the returned photo noted: the image depicts the catheter being used during a surgical procedure.The red luer adapter is disengaged from the extension tube.Without the physical device functional evaluation is precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the catheter's tip (tube head) had prolapsed during crrt (continuous renal replacement therapy).The catheter was not repaired and there was a leak at the luer connector.No cracks observed in the luer joint and iodophor was used as cleaning agent on the device.Tego was not utilized and the insertion site was treated prior to product placement.They had to removed the device to resolve the issue and the adapters were tighten by hands.The leak was located in the catheter fittings and extension tubes (arterial side).There was no other products being utilized with the device and the procedure was completed.There was no blood loss and no medical intervention done to the patient.There was no reported patient outcome.
|
| |
|
Search Alerts/Recalls
|