• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEX03 MEXICO-JUAREZ PRESOURCE CARDIAC CATH PACK (CBCHQ)840; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEX03 MEXICO-JUAREZ PRESOURCE CARDIAC CATH PACK (CBCHQ)840; GENERAL SURGERY TRAY (KIT) Back to Search Results
Model Number SAN11CBCHR
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
From the device history record, lot number 20190816-23-sh was manufactured on august 16, 2018.No exception was recorded in the device history record that could lead to the reported incident.No sample was available for evaluation.The average linting data is 0.202g / 10 pieces.The linting test data were within the acceptable range, therefore, the root cause could not be determined from this investigation.According to our supplier, or towels are made of cotton, so lint is born and inevitable.We are continuously working to better control the linting and have implemented the following: suctions machines have been installed in grey cloth rolling process, dyeing process and cutting process.The suction process was added before product's final folding, and workers do it according to standard operation procedure requirement.Linting test method and acceptable criteria was stipulated to see the suction results.(=0.38g/10 pieces).In the folding process, the supplier used one cloth pad under 100 pieces semi-finished products to avoid linting stuck onto the products during product's transfer.The complaint information was informed to the relevant sectors for their awareness.There is no action taken at this time, however, will continue to monitor the trend of this type of incident.
 
Event or Problem Description
Based on information received from the customer, reportedly some of the towels have a noticeable abundance of lint.Customer reported no injury to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIAC CATH PACK (CBCHQ)840
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
MEX03 MEXICO-JUAREZ PRESOURCE
c. arcadia 1580 , col. terraza
cd.juarez , chihuahua CP 32 599
MX  CP 32599
MDR Report Key9847318
Report Number1423537-2020-00437
Device Sequence Number16823702
Product Code LRO
UDI-Device Identifier10888439004267
UDI-Public10888439004267
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Model NumberSAN11CBCHR
Device Catalogue NumberSAN11CBCHR
Device Lot Number369759
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2020
Initial Report FDA Received Date03/18/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age73 YR
Patient Weight73
-
-