MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIG30AV
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a thoraco-lobectomy procedure, the surgeon tried to fire the device, but it couldn't fire.The device was not used in the patient.A cartridge was replaced then it worked, there was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that there were no abno rmalities that would have caused or contributed to the reported condition.The reload was loaded into a pmv representative signia handle for functional testing.The reload was cycled without hesitation or binding and was applied to test media.All staples were placed, and test media was cleanly transected.Functional testing confirmed the safety interlock feature successfully prevented the reload from cycling again.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key9847388
Report Number1219930-2020-01211
Device Sequence Number2533804
Product Code GDW
UDI-Device Identifier10884521543676
UDI-Public10884521543676
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberSIG30AV
Device Catalogue NumberSIG30AV
Device Lot NumberN9H0136Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2020
Initial Date Received by Manufacturer 03/03/2020
Supplement Date Received by Manufacturer04/16/2020
Initial Report FDA Received Date03/18/2020
Supplement Report FDA Received Date05/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-