| Model Number |
SIG30AV |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a thoraco-lobectomy procedure, the surgeon tried to fire the device, but it couldn't fire.The device was not used in the patient.A cartridge was replaced then it worked, there was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that there were no abno rmalities that would have caused or contributed to the reported condition.The reload was loaded into a pmv representative signia handle for functional testing.The reload was cycled without hesitation or binding and was applied to test media.All staples were placed, and test media was cleanly transected.Functional testing confirmed the safety interlock feature successfully prevented the reload from cycling again.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|