• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMERIEUX SA VIDAS MEASLES IGG; VIDAS® MEASLES IGG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMERIEUX SA VIDAS MEASLES IGG; VIDAS® MEASLES IGG Back to Search Results
Model Number 30219
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
A customer in (b)(6) notified biomérieux of obtaining false negative results in association with the vidas® measles igg 60tests (ref.30219, lot 1007322860).The customer stated seventeen patients were tested using vidas® measles igg 60tests lot 1007322860 and using the virotech test method.Seven (7) patients had no interpretation change in their results.Ten (10) patients obtained a negative result with the vidas® vidas® measles igg 60tests and a positive result using virotech.There is no indication or report from the laboratory that the discrepant results led to any adverse event related to any patient's state of health.Biomérieux will initiate an internal investigation.
 
Additional Manufacturer Narrative
This report was initially submitted following notification from a customer in croatia regarding false negative results in association with the vidas® measles igg 60tests (ref.30219, lot 1007322860).The customer stated seventeen (17) patients were tested using vidas® measles igg 60tests lot 1007322860 and using the virotech test method.Seven (7) patients had no interpretation change in their results.Ten (10) patients obtained a negative result with the vidas® vidas® measles igg 60tests and a positive result using virotech.A biomérieux internal investigation has been completed with the following results: customer's material: five (5) patients' samples with id 2-21656 - 2-21487, 2-21485, 2-21484 and 2-22860 were received.Complaint trending analysis: complaint analysis found one complaint for sensitivity issues and two complaints for discrepant results on vidas measles igg batch 1007322860/200325-0.There is neither capa nor non conformity linked to the object of this issue.Analysis of control charts: analysis of five (5) internal samples from activity panel was carried out on eight (8) lots of vidas measles igg including the customer lot 1007322860/200325-0.Vidas measles igg lot 1007322860/200325-0 had a trend to give results slightly lower compared to other lots but remained compliant to the release criteria of this vidas assay.**tests performed internally** - on internal samples five (5) positive samples with a positive target were selected according to their weak indexes and two (2) of them had a target just above the positive threshold.They were tested on retain kits of the vidas measles igg batch mentioned by the customer 1007322860/200325-0.The five (5) samples were within their acceptable range.There was no global evolution observed over time.On customer's samples: five (5) patients' samples were tested on four (4) different vidas measles igg batches including the batch mentioned by the customer.The negative results observed by the customer for samples 21484, 21485, 21656 and 22860 were reproduced on all vidas measles igg batches including the one mentioned by the customer.For sample 21487, there was a distribution of results in both side of the thresholds depending on the vidas measles igg batch used.According to the investigation conducted by the industrialization and support team, including tests performed on additional samples from internal biobank, there is no reconsideration of the between-lot variability of vidas measles igg assay ref.30219 which still be in accordance with the precision profile of vidas measles parameter.It is mentioned in the vidas measles igg assay ref.30219 in the results and interpretation section: the imprecision inherent in any method implies a lack of confidence in samples with test values very close to the thresholds.Consequently, an equivocal zone is established between the thresholds based on a statistical understanding of this imprecision.Conclusion: although negative results were reproduced on some samples (e.G customer's samples) and some differences of test values were observed, there is no reconsideration of vidas measles igg assay performances.The results observed on one patient's sample is due to its activity close to the threshold leading a distribution on both sides.According to all the data issued during the both investigations, there is no reconsideration of the vidas measles igg lot.There is no uprising trend of complaints on the lot mentioned by the customer.The vidas measles igg lot 1007322860/200325-0 was still within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDAS MEASLES IGG
Common Device Name
VIDAS® MEASLES IGG
Manufacturer (Section D)
BIOMERIEUX SA
376 chemin de l orme
marcy l etoile, 69280
FR  69280
MDR Report Key9847919
Report Number8020790-2020-00028
Device Sequence Number11481184
Product Code LJB
UDI-Device Identifier03573026064532
UDI-Public03573026064532
Combination Product (Y/N)N
Initial Reporter CountryHR
PMA/510(K) Number
C1 EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/25/2020
Device Model Number30219
Device Catalogue Number30219
Device Lot Number1007322860
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2020
Supplement Date Received by Manufacturer06/05/2020
Initial Report FDA Received Date03/18/2020
Supplement Report FDA Received Date06/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-