Model Number PM3210 |
Device Problems
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
Patient Problem
Dizziness (2194)
|
Event Date 02/15/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Related manufacturer reference number: 2017865-2020-02844.It was reported that the patient presented in clinic for routine follow-up experiencing dizziness.Upon interrogation, the pacemaker and right ventricular lead exhibited oversensing of noise.Far-r wave oversensing was also noted.No device intervention was performed.Patient condition was stable.
|
|
Manufacturer Narrative
|
The reported field event of over-sensing was not confirmed in the laboratory.The device was tested on the bench and using automated testing equipment, and no anomalies were found.
|
|
Event Description
|
Additional information received indicated that the patients' right ventricular (rv) lead was capped and replaced on (b)(6) 2021.The patient's pacemaker was also explanted and replaced.The patient was stable throughout.
|
|
Search Alerts/Recalls
|