| Model Number |
97725 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Energy Output Problem (1431)
|
| Health Effect - Clinical Codes |
Fall (1848); Device Overstimulation of Tissue (1991); Pain (1994); Hip Fracture (2349); Electric Shock (2554)
|
| Date of Event |
03/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received by a manufacture representative (rep) and consumer regarding a patient's implantable neurostimulator (ins).It was reported from the rep stating they were on the phone with the patient and coached them through changing group on the patient controller.They were trying to have them turn the stimulation off then lower ma before changing to the next group because they told the rep that the settings from day before on group a were too strong as the day progressed.After multiple attempts to press the white on/off button on top of the controller, they were unable to press the button to get the screen to turn off.They practiced this step several times the day before in pacu (post-anethesia care unit) and it was a little difficult to them due to age, strength, and dexterity.The patient was warned that they will likely have to decrease as soon as switching to group c and to make sure they were sitting down.When they changed, the patient got extreme shock and started turning down, but they fell to the floor before they could get low enough.The rep spoke to the patient the next day and the patient reported their leg and groin hurt and they were going to urgent care or emergency room to get checked.They were diagnosed with a fractured art hip.No further complications reported/anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received from the healthcare professional (hcp) via a rep and it was reported that there was no device issue.Patient's group c ma was set appropriately at the facility.When the patient switched to it a home, the stimulation felt stronger.No further complications reported/anticipated.
|
| |
|