• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. BASE BAR ASSM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. BASE BAR ASSM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 210060
Medical Device Problem Codes Degraded (1153); Material Frayed (1262)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event or Problem Description
F-bar is beginning to splinter.Case type: tka.
 
Event or Problem Description
F-bar is beginning to splinter.Case type: tka.
 
Additional Manufacturer Narrative
Update to b2 and d2.Reported issue - f-bar is beginning to splinter.Product inspection - visual inspection of catalog #210060, lot/serial (b)(6) confirms the base bar has splintered along the side of the bar.See image attached.Product history review ¿ review of the device history records for catalog #210060, lot/serial (b)(6) indicates (b)(4) devices were manufactured and all (b)(4) devices were accepted into final stock on 12/27/2018 with no reported discrepancies.Complaint history review ¿ a review of complaints in catsweb and trackwise related to p/n 210060, lot #615603 shows 0 additional complaints related to the failure in this investigation.Conclusion ¿ the event was confirmed.There was splintering noticed along the base bar.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE BAR ASSM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key9850208
Report Number3005985723-2020-00166
Device Sequence Number10953853
Product Code OLO
UDI-Device Identifier00848486031855
UDI-Public00848486031855
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number210060
Device Catalogue Number210060
Device Lot Number615603
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/24/2020
Supplement Date Received by Manufacturer06/05/2020
Initial Report FDA Received Date03/18/2020
Supplement Report FDA Received Date06/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
-
-