• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VNMC3428C207TU
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 02/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant navion stent graft was implanted in the endovascular treatment of a 200mm aortic thoracic dissection.It was reported five months later a reterograde type a dissection occurred and the patient received treatment two days later.The details of the treatment that the patient received are unknown.As per the physician the cause of the event is unknown.No additional clinical sequelae were provided and the patient is fine.
 
Additional Manufacturer Narrative
Film evaluation summary: the reported type a dissection was confirmed by the films provided; however the cause of the event could not be determined.Additional ct¿s or angiograms to allow more visualisation of the type a dissection were not provided for a more thorough analysis.From the images received it is difficult to ascertain the exact location of the new dissection in relation to the area in which the valiant navion was implanted in.It is unclear if the reported type a dissection was possibly due to a patient issue and the previous type b dissection, or possibly caused by an unknown interaction with the implanted stent graft.Analysis of the returned films did not reveal any anatomical characteristics or device issues that could explain the reported dissection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key9852165
Report Number9612164-2020-01219
Device Sequence Number16840898
Product Code MIH
UDI-Device Identifier00763000101237
UDI-Public00763000101237
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/15/2020
Device Model NumberVNMC3428C207TU
Device Catalogue NumberVNMC3428C207TU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2020
Initial Date Received by Manufacturer 02/20/2020
Supplement Date Received by Manufacturer03/27/2020
Initial Report FDA Received Date03/19/2020
Supplement Report FDA Received Date03/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age85 YR
-
-