B. BRAUN MEDICAL INC. EXTENSION SET; SET, ADMINISTRATION, INTRA Back to Search Results
Model Number 470195
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Chemical Exposure (2570); No Known Impact Or Consequence To Patient (2692)
Date of Event 02/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The device involved has been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: customer reports two (2) chemotherapy incidents.Event 2: it was reported that the nurse noted that the product was leaking.The line was paused and changed.Chemotherapy drug involved was not provided.No injury reported.Event 1: was reported via mfr report number: 2523676-2020-00080.It was reported that during chemotherapy of busulfan, the product leaked.The lower y-site leaked.Chemo leaked on patients arm, shirt and bed.No injury reported.
 
Additional Manufacturer Narrative
Event 2: this report has been identified as b.Braun medical internal report number (b)(4).Three (3) samples in total were returned for evaluation.One sample was used with packaging, two samples were unused with packaging.The used sample was visually inspected per specification and it was noted the lower caresite lure access device (y-site) had a crack on the threads.The used sample was also tested for leakage per specifications with failing results.The remaining two unused samples were tested for leakage per specifications with passing results.Based on the evaluation results the reported defect was confirmed.The exact root cause could not be determined at this time.However, incidents of cracked/leaking valves can be caused by several factors, including but not limited to the product being subjected to aggressive solvents/drugs, excessive mechanical stresses (over-tightening or frequent set manipulation), or other various unforeseen circumstances during the clinical application.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.In addition, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.Furthermore, the house retain samples were tested per specifications and no defects were observed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown, pa
MDR Report Key9852794
Report Number2523676-2020-00081
Device Sequence Number2533441
Product Code FPA
UDI-Device Identifier04046964823428
UDI-Public04046964823428
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K153293
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number470195
Device Catalogue Number470195
Device Lot Number0061700009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2020
Initial Date Received by Manufacturer 03/05/2020
Supplement Date Received by Manufacturer03/05/2020
Initial Report FDA Received Date03/19/2020
Supplement Report FDA Received Date04/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-