| Model Number |
70103 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a low recovery (30%) and purity (60%) for a cd34 separation in an autologous setting for a neuroblastoma patient.Nothing out of the ordinary was observed during the separation run.Also the volume of the target fraction was as expected.The number of target cells in the starting material was low with 1.8x106 cd34 positive cells.The patient was not treated with the product.An investigation concluded that there is no hint for a product malfunction or user error but that rather the composition and characteristics of this starting material lead to the dissatisfying results.The frequency of cd34+ cells in the starting material was very low and the material was derived from a pre-treated patient.
|
| |
|