| Model Number |
SCX-310 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/18/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: analysis of the returned device identified: weight to the spec, opening curved on anterior, red particles in the gel and creases fold.A visual and microscopic analysis was performed which identified: striated opening on anterior.Base on the device analysis the final assessment is: a striated opening on anterior assessed as surgical damage consist in the use of some surgical tool.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device has been explanted.
|
| |
|