• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATUN FEM SLV M/L 35MM HALF POR; KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATUN FEM SLV M/L 35MM HALF POR; KNEE FEMORAL ACCESSORY Back to Search Results
Model Number 1511-01-103
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); No Code Available (3191)
Date of Event 03/02/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision of components due to infection, all components removed from previous revision case done on the (b)(6) 2020 which was the 2nd stage revision, 1st stage revision completed on the (b)(6) 2019.Primary surgeon //name: (b)(6), primary surgery facility: (b)(6), this is the 3rd stage revision.
 
Additional Manufacturer Narrative
H10 additional narrative: product complaint #
=
> pc-(b)(4).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Device history lot
=
> null.Device history batch
=
> null.Device history review
=
> null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUN FEM SLV M/L 35MM HALF POR
Common Device Name
KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9853001
Report Number1818910-2020-08580
Device Sequence Number3332377
Product Code NJL
UDI-Device Identifier10603295436508
UDI-Public10603295436508
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1511-01-103
Device Catalogue Number151101103
Device Lot NumberHW5278
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/03/2020
Supplement Date Received by Manufacturer03/23/2020
04/16/2020
Initial Report FDA Received Date03/19/2020
Supplement Report FDA Received Date03/30/2020
04/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS FEMORAL LT SZ 6 CEM; ATTUNE CRS RP INSRT SZ 6 6MM; ATTUNE DIST FEM AUG SZ 6 12MM; ATTUNE DIST FEM AUG SZ 6 4MM; ATTUNE POS FEM AUG SZ 6 8MM; ATTUNE POS FEM AUG SZ 6 8MM; ATUN FEM SLV M/L 35MM HALF POR; ATUN TIB SLV M/L 29MM HALF POR; ATUNE PRESSFIT STR STEM14X60MM; ATUNE PRESSFIT STR STEM16X60MM; ATUNE REV RP TIB BASE SZ 5 CEM; COMPETITOR CEMENT.; ATTUNE CRS FEMORAL LT SZ 6 CEM; ATTUNE CRS RP INSRT SZ 6 6MM; ATTUNE DIST FEM AUG SZ 6 12MM; ATTUNE DIST FEM AUG SZ 6 4MM; ATTUNE POS FEM AUG SZ 6 8MM; ATTUNE POS FEM AUG SZ 6 8MM; ATUN FEM SLV M/L 35MM HALF POR; ATUN TIB SLV M/L 29MM HALF POR; ATUNE PRESSFIT STR STEM14X60MM; ATUNE PRESSFIT STR STEM16X60MM; ATUNE REV RP TIB BASE SZ 5 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-