• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR RX SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR RX SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550350-33
Medical Device Problem Code Expiration Date Error (2528)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Event or Problem Description
It was reported that a 3.5x33mm xience sierra stent delivery system was labeled with a shelf life of three years when it should have been labeled with a shelf life of one year.The device was not used and there was no patient involvement.No additional information was provided.
 
Additional Manufacturer Narrative
On march 19, 2020, abbott vascular determined that a field safety corrective action is required regarding one lot of the xience sierra rx everolimus eluting coronary stent system.Devices from this lot were shipped with a 36-month labeled expiration date instead of the 12-month expiration date that is currently approved in the united states.The label indicates an expiration date of march 21, 2022, while the fda approved expiration date is march 31, 2020.This action does not affect patients having successfully undergone procedures using these devices.The customer reported the device is retained by hospital and not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.H7 - if remedial action initiated updated from "none selected" to "notification".
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified the nonconformity issued to the reported lot identifying the reported issue of expiration date error addressed in this investigation.A review of the complaint history identified no other incidents from this lot.The investigation determined the reported expiration date error appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX SIERRA
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9856203
Report Number2024168-2020-02654
Device Sequence Number3307478
Product Code NIQ
UDI-Device Identifier08717648227455
UDI-Public08717648227455
Combination Product (Y/N)N
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number1550350-33
Device Catalogue Number1550350-33
Device Lot Number903224A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/25/2020
Supplement Date Received by Manufacturer03/19/2020
03/26/2020
Initial Report FDA Received Date03/19/2020
Supplement Report FDA Received Date03/24/2020
04/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-