• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE TEMP SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE TEMP SENSING FOLEY CATHETER Back to Search Results
Model Number 119208M
Medical Device Problem Code Misassembled (1398)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the 8 fr temp-sensing foley was found to be not patent after attempted insertion in a patient.Per additional information received, the foley did not drain.When the urologist came to assist, he examined the foley after the removal and there was no hole in the end of the foley for the urine to drain out.
 
Event or Problem Description
It was reported that the 8fr temp-sensing foley was found to be not patent after attempted insertion in a patient.Per additional information received, the foley did not drain.When the urologist came to assist, he examined the foley after the removal and there was no hole in the end of the foley for the urine to drain out.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A potential failure mode could be ¿no eyes¿ with a potential root cause of "inadequate process".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿additional safety instructions include the following: 1.The foley catheter with temperature sensor should not be connected to the temperature monitoring equipment during the mri procedure.2.If the foley catheter with temperature sensor has a removable catheter connector cable, it should be disconnected prior to the mri procedure.3.Remove all electrically conductive material from the bore of the mr system that is not required for the procedure (i.E., unused surface coils, cables, etc.).4.Keep electrically conductive material that must remain in the bore of the mr system from directly contacting the patient by placing thermal and/or electrical insulation (including air) between the conductive material and the patient.5.Position the foley catheter with a temperature sensor in a straight configuration down the center of the patient table to prevent cross points and conductive coils or loops.6.The wire and connector of the foley catheter with temperature sensor should not be in contact with the patient during the mri procedure.Position the device, accordingly.7.Mr imaging should be performed using an mr system with static magnetic strength of 1.5-tesla or 3-tesla, only.8.At 1.5-tesla, the mr system whole body averaged sar should not exceed 3.5- w/kg for 15-min.Of scanning.9.At 3-tesla, the mr system reported whole body averaged sar should not exceed 3-w/kg for 15-min of scanning." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE TEMP SENSING FOLEY CATHETER
Common Device Name
TEMP SENSING FOLEY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9856402
Report Number1018233-2020-01959
Device Sequence Number17003237
Product Code EZL
UDI-Device Identifier00801741046261
UDI-Public(01)00801741046261
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/31/2024
Device Model Number119208M
Device Catalogue Number119208M
Device Lot NumberNGDU1415
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/27/2020
Supplement Date Received by Manufacturer04/17/2020
Initial Report FDA Received Date03/19/2020
Supplement Report FDA Received Date04/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-