The device history record (dhr) was not performed during this investigation as the lot number was not received with the complaint.All dhr¿s are reviewed and approved by quality prior to the release of the product.It is also important to note that a series of tests are performed during every lot of production.More specifically, inspections are performed to test the seal strength for both the inner seal, where the product would activate and the outer seal.During the test, the pouch is first activated, the inner seal is broken.The pouch is then placed between two plates.The pouch is compressed until the pressure required per the procedure is reached, 350 lbs.Minimum.The pouch is left to dwell in the machine for 20 seconds, using a calibrated stopwatch to time it.After the dwell time the pouch is removed and visually inspected for leaks or spreading of the seam.A pouch with any deficiency would be rejected at this time and the machine adjusted.If the pouch passes the dwell test it is returned to the machine and the pouch is the compressed until the pouch breaks or reaches over 1000 pounds.The customer returned samples in the form of 20 heel warmers.A visual inspection was completed on each sample.The results were that 19 samples showed a hole in the top left (as you look at the back of the pouch) or full top of the clear side of the pouch and 1 sample showed pre-activation bit it did not have a hole in the heel warmer pouch.While a definitive root cause could not be determined based solely on the returned sample, a likely contributor for this issue is as the vertical sealer bar is sealing the top of two consecutive pouches, there could be a localized weak region just prior to the midpoint of the vertical sealer bar causing an incomplete seal.Additionally, there could have been some type of buildup on the sealer bars that would cause a localized weak spot.Relative to the issue reported by the customer, pressure was applied by the clinician and this area of the pouch broke open.The results of the manufacturing facility investigation were able to identify a localized area of the pouch that was the likely area where the pouch contents leaked out of.A review of maintenance activity noted that the sealer bars were dirty and worn and were therefore cleaned and replaced.The operator¿s preventive maintenance (pm) guide has been updated to include a visual inspection and cleaning of all sealer bars and dies every month during regular pm¿s.
|