| Model Number |
PROPAQ MD |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device displayed a "defib pacer device failure" message.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, power cycling and full defib/pacer functionality testing without duplicating the report.The auxillary power supply was not returned as part of this investigation.The customer was advised to inspect their auxillary power supply as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|