• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS INSERT SZ 5-6 11MM; PRSTHSS,KN,PTLLFMRTBL,SM-CNSTRND,CMNTD,PLYMR/MTL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS INSERT SZ 5-6 11MM; PRSTHSS,KN,PTLLFMRTBL,SM-CNSTRND,CMNTD,PLYMR/MTL Back to Search Results
Model Number 71420832
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Unspecified Infection (1930); Injury (2348)
Date of Event 11/07/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a revision surgery was performed due to infection.The entire system was replaced.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.Only the tibial insert was returned for evaluation.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage and wear from implantation and extraction.The device was manufactured in 2007.The medical investigation concluded that, without the requested clinical information, a thorough medical investigation of the revision due a reported infection cannot be rendered.Should any additional medical information be provided this complaint will be re-assessed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the sterilization records revealed the batch was sterilized within normal parameters.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Infection is a complication that can be associated with any surgery.Some potential probable causes could include contamination, patient reaction or a post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS INSERT SZ 5-6 11MM
Common Device Name
PRSTHSS,KN,PTLLFMRTBL,SM-CNSTRND,CMNTD,PLYMR/MTL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9856832
Report Number1020279-2020-00936
Device Sequence Number2533393
Product Code JWH
UDI-Device Identifier03596010209658
UDI-Public03596010209658
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/20/2017
Device Model Number71420832
Device Catalogue Number71420832
Device Lot Number07FM18492
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/07/2019
Supplement Date Received by Manufacturer06/11/2020
Initial Report FDA Received Date03/19/2020
Supplement Report FDA Received Date06/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-