| Model Number |
NSLG2C45 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # t94e62.Investigation summary: the device was returned with the bottom portion of the i blade out of the lower jaw channel; the lower jaw channel was damaged.The device was connected to the generator and it was recognized.Because of the condition of the device not all functional testing could be performed with the generator.Therefore we were unable to investigate further the reported event.The jaw was unable to cycle open and close.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.A probable cause of the damage to the jaw could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic duodenal switch the device failed to activate/malfunctioned after initial use.A second device was opened and had the same issue.A third device was used to complete the case with no patient consequences.
|
| |
|