| Model Number |
00884450310151 |
| Medical Device Problem Codes |
Material Separation (1562); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a thrombectomy procedure using a balloon to clear the excessive thrombus from the venous output component [voc], the distal ring on the voc device detached and migrated to the patient's lung.A new voc was placed for continued treatment.The patient tolerated the procedure well and remained asymptomatic.No additional procedures have been scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.The complaint is confirmed.The root cause is attributed to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Corrective action for this issue has been implemented via re-design of the product.
|
| |
|