SMITH & NEPHEW, INC. ACCESSORY TRACTION; ACCESSORIES, OPERATING-ROOM, TABLE (KIT) Back to Search Results
Model Number 7210577
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/27/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a shoulder surgery, the system holder is not getting locked.It is loose and cannot handle the shoulder.It is unknown how the procedure was completed however no significant delay and no patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint facility for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be re-opened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESSORY TRACTION
Common Device Name
ACCESSORIES, OPERATING-ROOM, TABLE (KIT)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key9857561
Report Number1643264-2020-00157
Device Sequence Number2611459
Product Code FWZ
UDI-Device Identifier03596010511256
UDI-Public03596010511256
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7210577
Device Catalogue Number7210577
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2021
Initial Date Received by Manufacturer 02/27/2020
Supplement Date Received by Manufacturer04/21/2020
04/20/2021
Initial Report FDA Received Date03/19/2020
Supplement Report FDA Received Date04/26/2020
07/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-