Model Number V60
Device Problem Decrease in Pressure (1490)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 19mar2020.
 
Event Description
The customer reported that the ventilator alarmed due to oxygen pressure being low.The device was being used for treatment when the reported event occurred; however, there was no patient harm.The manufacturer's international service technician evaluated the device and confirmed the reported problem.The service technician replaced the data acquisition board and the reported problem was resolved.
 
Manufacturer Narrative
G4: 01apr2020.B4: 06apr2020.H11: b5 correction to problem statement: the customer reported that the ventilator produced the proximal pressure sensor autozero diagnostic failure.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9857852
MDR Text Key186683840
Report Number2031642-2020-00973
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public(01)00884838025776
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/02/2020
Initial Date FDA Received03/19/2020
Supplement Dates Manufacturer Received03/02/2020
Supplement Dates FDA Received04/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-