Product event summary: the transseptal needle, ep003994s with lot number 212632798, was returned and analyzed.There were no anomalies found.The reported skiving issue was not confirmed; however, it is plausible that skiving occurred as there is a compatibility issue with the needle and a competitor sheath.The cause of the issue was not established.If information is provided in the future, a supplemental report will be issued.
Event or Problem Description
It was reported that during a radiofrequency (rf) ablation procedure, the transseptal needle was scratching the inside of the dilator and could not be advanced out of the dilator.The needle and sheath were replaced.No patient complications have been reported as a result of this event.The case was completed and no debris or particles were noticed.