MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 5484902
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/19/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent percutaneous pedicle screw fixation (pps) surgery at t12/l1/l3 due to l2 burst fracture.Intra-op, on the right side of t12, the reducer was used to drop the rod down.When the reducer had been dropped down with the reduction nut driver and the egg handle, the nut part of the reducer broke.No patient complications were reported.It was said that since it became unable to use the reducer, the extender was removed and the rod was pressed down using the counter torque, and final tightening could be performed.
 
Additional Manufacturer Narrative
Additional information: d4, d10, h3, h4, h6 h6: product analysis results: visual and optical inspection confirmed the retaining ring that holds the reduction nut to the reduction body has failed.There are witness marks on the ear of the shaft indicating that there was some force applied to the reducer.It appears the ring was overloaded from excessive force during use.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key9858661
Report Number1030489-2020-00337
Device Sequence Number16971073
Product Code NKB
UDI-Device Identifier00643169177024
UDI-Public00643169177024
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K122862
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5484902
Device Lot NumberEM15F040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2020
Initial Date Received by Manufacturer 02/19/2020
Supplement Date Received by Manufacturer05/26/2020
Initial Report FDA Received Date03/20/2020
Supplement Report FDA Received Date06/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
-
-