Keel punches and handles disengage leaving the punches in the bone unable to retrieve was surgery delayed due to the reported event? yes, if yes, number of minutes: 1, action taken when event occurred? used osteotome to retrieve, was procedure successfully completed? yes, were fragments generated? no, if yes, were they removed easily without additional intervention? --> unknown, patient status/ outcome / consequences: no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: no, is the patient part of a clinical study: unknown, (b)(4), device property of: none, device in possession of: none, (b)(4), device property of: none, device in possession of: none, (b)(4), device property of: none, device in possession of: none, by checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.True.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned, thus the reported event could not be confirmed.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|