• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Melted (1385)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/22/2020
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. The standard for a (b)(4) plug top and socket, which is used in (b)(4), is 13a (ampère) and heater cooler is rated 16a. This caused the heating which lead to burned plug top and alarm in the electrical panel of the building. A 16a socket will be used to solve the issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.

 
Event Description

Livanova (b)(4) received a report that the power cable and the plug top of a heater-cooler system 3t burned due to overheating during procedure. There was no report of patient injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM 80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9860291
MDR Text Key197145973
Report Number9611109-2020-00209
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeQA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 03/19/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received03/20/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/22/2020
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured02/02/2010
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-