| Model Number |
TMMB10 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Patient identifier unknown / not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant at tooth site #47 fractured above the trabecular metal portion and it was completely removed.
|
| |
|
Additional Manufacturer Narrative
|
|
One implant was returned for investigation.Inspection of the as returned product confirmed that the device is fractured.A device history review was performed and no related nonconformance¿s were noted.A device history review was performed and no related nonconformance¿s were noted.A complaint history search was performed using our complaint handling system and there were no additional related complaints against this lot.Appropriate documentation was reviewed.The alleged device malfunction was confirmed.A root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|