COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1737
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during procedure, the hand piece's coagulation was inadequate but cut was okay.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The reported condition was not confirmed.Functional testing was performed with the returned device on porcine kidney tissue.Multiple seals on various size vessels were made and a full thermal effect was visually verified.The device was activated while pressing the button and with the rotation knob in various positions to detect any problematic activation issues.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic gastric sleeve, the hand piece's coagulation was inadequate but cut was okay.Another hand piece was used without issues.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key9860688
Report Number1717344-2020-00317
Device Sequence Number11589979
Product Code GEI
UDI-Device Identifier10884521181571
UDI-Public10884521181571
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K133338
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/04/2024
Device Model NumberLF1737
Device Catalogue NumberLF1737
Device Lot Number90650112V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2020
Initial Date Received by Manufacturer 03/03/2020
Supplement Date Received by Manufacturer03/21/2020
Initial Report FDA Received Date03/20/2020
Supplement Report FDA Received Date04/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-