| Model Number |
06002-CP-110 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Donor was monitored and was stable upon release.A field service engineer was dispatched and inspected the pcs2 machine.The engineer realigned centrifuge lid and tightened loose hardware.The blood pump motor was also replaced.Calibrations checked and adjusted as required.Unit meets manufacturer specifications.Photo analysis of the bowl shows a bowl full of dark colored red blood cells.No evidence of damaged red blood cells are visible.No defects that could possibly create hemolysis were identified (no folds, molding defects, sharp or rough edges, foreign objects).
|
| |
|
Event or Problem Description
|
|
On february 21, 2020, haemonetics was informed by the customer of an adverse event and device malfunction during a plasmapheresis procedure.There was a possibility of the donor receiving hemolized red blood cells.
|
| |
|