After review of medical records patient was revised to addressed failed metal on metal right hip arthroplasty with metallosis and abductor muscle necrosis.Operative notes indicated elevated metal ion level, corrosion, discomfort and pseudotumor formation.Upon entering into the joint, the metallic fluid was under tension and had evidence of metal debris.There was evidence of corrosion at the taper of the liner acetabular component interface with pseudomembrane formation on the undersurface of the acetabular liner.Doi: (b)(6) 2009; dor: (b)(6) 2019 right hip.
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|