WRIGHTS LANE SYNTHES USA PRODUCTS LLC SYNFRAME HOLDING BASE INSULATED F/OR TAB; MISC ORTHO SURGICAL INSTR Back to Search Results
Model Number 387.346
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date the 2 synframe table clamps keep getting stuck on the operation table rails and are very difficult to unscrew at the end of surgery.Or staff had to use vise grips to unscrew the clamps.It¿s happened several times.It was unknown if there was surgical delay.There was no patient consequence.This complaint involves two (2) devices.This is 1 of 2 for report (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary.Investigation flow: functional/device interaction.Visual inspection: the synframe holding base insulated f/or tab (p/n: 387.346, lot number: 1136249) was received at us cq.Visual inspection of the complaint device showed the large black screw handle was hard to turn and jammed when fully loosened.Functional test: a functional assessment was performed with the complaint device.One of the screw handles was hard to turn and jammed when loosened.The condition can be replicated.Dimensional inspection: a dimensional inspection was not performed due to internal components being inaccessible.Document/specification review: no design issues or discrepancies were identified.Complaint confirmed? yes.Investigation conclusion: this complaint is confirmed as the device is difficult to use and jams when loosened.No root cause could definitively be determined for the reported complaint condition.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot.Part: 387.346, lot: 1136249, manufacturing site: haegendorf, release to warehouse date: aug.27, 2002.Expiry date: sep.01, 2017.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNFRAME HOLDING BASE INSULATED F/OR TAB
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9862354
Report Number2939274-2020-01429
Device Sequence Number2611014
Product Code LXH
UDI-Device Identifier10705034771120
UDI-Public(01)10705034771120
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2017
Device Model Number387.346
Device Catalogue Number387.346
Device Lot Number4473795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2020
Initial Date Received by Manufacturer 02/24/2020
Supplement Date Received by Manufacturer03/31/2020
04/23/2020
Initial Report FDA Received Date03/20/2020
Supplement Report FDA Received Date04/22/2020
05/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADAPT F/JACKSON OR TABLE L195 F/387.346+; ADAPT F/JACKSON OR TABLE L195 F/387.346+
-
-