• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETER; TIEMANN CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® URETHRAL CATHETER; TIEMANN CATHETER Back to Search Results
Model Number 120618
Health Effect - Clinical Codes Urinary Retention (2119); No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the patient stated he had several episodes of not being able to insert the catheters.It was unknown if the patient was able to void.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential failure mode could be ¿difficult insertion¿ with a potential root cause of ¿outside diameter is too large¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "warnings: ¿ on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.¿ visually inspect the product for any imperfections or surface deterioration prior to use.¿ reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the patient stated he has had several episodes of not being able to insert the catheters.It is unknown if the patient was able to void.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URETHRAL CATHETER
Common Device Name
TIEMANN CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9862380
Report Number1018233-2020-01991
Device Sequence Number11546095
Product Code KOD
UDI-Device Identifier00801741029981
UDI-Public(01)00801741029981
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number120618
Device Catalogue Number120618
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/27/2020
Supplement Date Received by Manufacturer04/10/2020
Initial Report FDA Received Date03/20/2020
Supplement Report FDA Received Date04/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-