| Model Number |
72202588 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a hip procedure when the suture cutter cuts it leaves black dust.No delay and a back up was available to complete the procedure.No patient injury was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: one 72202588 xl suture cutter was returned for evaluation.Visual assessment confirms complaint.The information provided states: ¿during a hip procedure when the suture cutter cuts it leaves black dust¿.Per the device ifu ¿as with any surgical instrument, careful attention should be exercised to ensure that excessive force is not placed on the instrument.Excessive force can result in instrument failure¿.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.A review of complaints and manufacturing batch records was preformed, no other complaints of this failure was found.Further investigation is not warranted at this time.
|
| |
|