MAUDE Adverse Event Report: MEDTRONIC MEXICO KYPHON EXPRESS INFLATABLE BONE TAMP; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO KYPHON EXPRESS INFLATABLE BONE TAMP; ARTHROSCOPE Back to Search Results
Catalog Number K15B
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient presented with l1 vertebral compression fracture; and underwent l1 balloon kyphoplasty.Intra-op, while re-introducing the inner stylet into the bone tamp, the bone tamp burst inside the patient.The stylet pierced the inflated balloon.The procedure was then completed successfully with a new product.No patient complications were reported as a result of this event.
 
Additional Manufacturer Narrative
Product analysis: visual and functional inspection revealed the balloon has been punctured/ruptured.The rupture is a pinhole near the proximal peak of the balloon.This rupture is consistent with the balloon coming in contact with something sharp like bone spurs when the balloon was inflated in the vertebral body.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: b5 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
T8 radio-frequency ablation and kyphoplasty was being performed.The inflatable bone tamp burst at 300 psi.The procedure was delayed for approximately 10 minutes as the physician was trying to flush contrast out of vertebral body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON EXPRESS INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key9862812
Report Number9612164-2020-01266
Device Sequence Number3307921
Product Code HRX
UDI-Device Identifier00763000025625
UDI-Public00763000025625
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/26/2021
Device Catalogue NumberK15B
Device Lot Number218323315
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2020
Initial Date Received by Manufacturer 02/20/2020
Supplement Date Received by Manufacturer03/24/2020
05/15/2020
Initial Report FDA Received Date03/20/2020
Supplement Report FDA Received Date04/22/2020
06/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age84 YR
Patient Weight66
-
-