Catalog Number 2B8071
Device Problems Break (1069); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 02/24/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a vial-mate adapter experience coring incident while reconstituting vancomycin 1.5g/vial with a normal saline 500ml bag.This was identified after reconstitution, during setup and preparation.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added: the device was received for evaluation.Visual inspection was performed using the naked eye which observed that the device was not in the activated position and was not docked to a vial.Additional inspection observed a particulate matter in the attached drained solution bag.The particulate matter was identified to be a stopper material from the vial.The reported condition was verified.The cause of the condition was potentially due to the user misuse causing stopper material to dislodge into the fluid path.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIAL-MATE RECONSTITUTION DEVICE
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CLEVELAND
911 highway 61 north
po box 1058
cleveland MS 38732
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9862953
MDR Text Key184422176
Report Number1416980-2020-01612
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142600
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 04/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2B8071
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/24/2020
Initial Date FDA Received03/20/2020
Supplement Dates Manufacturer Received04/13/2020
Supplement Dates FDA Received04/20/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-