Catalog Number 2B8071 |
Device Problems
Break (1069); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a vial-mate adapter experience coring incident while reconstituting vancomycin 1.5g/vial with a normal saline 500ml bag.This was identified after reconstitution, during setup and preparation.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information was added: the device was received for evaluation.Visual inspection was performed using the naked eye which observed that the device was not in the activated position and was not docked to a vial.Additional inspection observed a particulate matter in the attached drained solution bag.The particulate matter was identified to be a stopper material from the vial.The reported condition was verified.The cause of the condition was potentially due to the user misuse causing stopper material to dislodge into the fluid path.Should additional relevant information become available, a supplemental report will be submitted.
|
|