| Model Number |
N/A |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial hip procedure the surgical team went to open a m/l taper stem, and there was no glue on the plastic container sealing the paper to the plastic.The paper was not tightly sealed to the plastic holding containing the implant.It was decided not to use the implant due to uncertainty of sterility.No adverse events have been reported as a result of this malfunction.Attempts were made to obtain additional information; however, none was available.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event was confirmed by visual examination of the device/packaging.Device history record (dhr) was reviewed and no discrepancies were found.The root cause of the reported issue is attributed to a manufacturing error.Visual evaluation of the returned product confirmed that no glue was applied to the sealing flange of the inner sterile cavity and the inner lid was not attached to the inner cavity.The sterility is compromised.The likely condition of the part when it left zimmer biomet control is considered non-conforming based on the evaluation of the returned product.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|