| Model Number |
26270 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned with the stent separated from the delivery system.The stent was not returned for analysis.A visual and microscopic examination identified no damage or issues with the stent cups or the stent holder.The stent impression was evident on the stent holder.A visual and tactile examination found the outer to be severely kinked at more than one location.This type of damage is consistent with excessive force being applied to the device.A visual examination of the inner shaft found it to be completely detached from the metal shaft at its bond.This type of damage is consistent with excessive tensile force being applied to the device.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2020.A 10x69x75/ iliac 6f unistep plus wallstent rp endoprosthesis was used with no known issues reported.However, returned device analysis revealed stent detachment and shaft break.
|
| |
|