| Model Number |
71411-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
A health care professional reported a high reading from an adc hospital blood glucose meter.The health care professional reported a high reading of 157 mg/dl which was higher than a lab reading of 61 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specifications.Dhrs for the precision test strips were reviewed and the dhrs showed the precision test strips passed all tests prior to release.Dhrs for the precision pro meter were reviewed and the dhrs showed the precision pro meter passed all tests prior to release.Retain testing was performed for precision strips and all units performed in the specification and passed.If the product is returned, a physical investigation will be performed and a follow-up report submitted.
|
| |
|
Event or Problem Description
|
|
A health care professional reported a high reading from an adc hospital blood glucose meter.The health care professional reported a high reading of 157 mg/dl which was higher than a lab reading of 61 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported meter has been returned and investigated with retained test strips.Visual inspection was performed on the returned meter and no issues were observed.Control solution testing has been performed and no issues were observed.All results were within range specification.No malfunction or product deficiency was identified.
|
| |
|
Event or Problem Description
|
|
A health care professional reported a high reading from an adc hospital blood glucose meter.The health care professional reported a high reading of 157 mg/dl which was higher than a lab reading of 61 mg/dl.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|