| Model Number |
8900-0400 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the associated device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another set of pads to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This report was inadvertently submitted as a duplicate.The reported event was originally reported under medwatch 1220908-2020-00688.Evaluation results were: the device was returned to zoll medical canada and the customer's report was observed during review of the device activity logs.However, the device was put through extensive testing including full functionality testing, stress testing and ecg signal testing without duplicating a malfunction to the device.Visual inspection of the electrode pads returned found that one of the high voltage wires appeared to be cut.It is not known if the damage occurred during use or after.Review of the logs does appear to show the device worked flawlessly upon replacement of the electrodes for the remainder of the case.No trend is associated with reports of this type.
|
| |
|