• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 4.0MM FEMALE HEX DRIVER; EXTREMITY INSTRUMENTS : SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 4.0MM FEMALE HEX DRIVER; EXTREMITY INSTRUMENTS : SCREWDRIVERS Back to Search Results
Model Number 2100-70-155
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4).Examination of the returned device confirmed the reported event.The damage is consistent with device use from normal use and servicing and the investigation did not establish a need for corrective action.Examination of the device found that the inside hex driver feature exhibits a crack, wear and stripping.There are cracks in the driver feature walls.There is no breakage or missing material.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the 4.0mm green driver was ¿stripped¿ according to surgeon customer.No surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported event.The damage is consistent with device use from normal use and servicing and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.0MM FEMALE HEX DRIVER
Common Device Name
EXTREMITY INSTRUMENTS : SCREWDRIVERS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9866273
Report Number1818910-2020-08860
Device Sequence Number19629079
Product Code HXX
UDI-Device Identifier10603295085959
UDI-Public10603295085959
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2100-70-155
Device Catalogue Number210070155
Device Lot NumberPG217705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2020
Initial Date Received by Manufacturer 03/04/2020
Supplement Date Received by Manufacturer03/24/2020
Initial Report FDA Received Date03/23/2020
Supplement Report FDA Received Date03/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Patient Weight82
-
-